Technical Aspects of the Quality Management System (QMS) for the Radiation Dosimetry Laboratory at ELI-NP Facility
Abstract
The individual and
environmental passive monitoring program of the Extreme Light Infrastructure -
Nuclear Physics (ELI-NP) will use a combination of Optically Stimulated
Luminescence (OSL) and Solid State Nuclear Track Detector – (SSNTD) detectors.
The OSL dosimeters are dedicated for X-rays/gamma radiation monitoring and will
contain BeO detectors which provide the personal dose equivalent and the
ambient dose equivalent or the directional dose equivalent according to SR EN
62387-1:2012. The SSNTDs are from poly-allyl diglycol carbonate (CR39), whose
performance and test requirements are conform to the criteria of ISO
21909-1:2015 and ISO 8529-3:1998 standards. All the activities of the Radiation
Dosimetry Laboratory from ELI-NP will comply the requirements of the ISO/IEC
17025:2005 standard. This paper presents the management of the technical
aspects of Quality Management System (QMS) implemented in the Radiation
Dosimetry Laboratory ELI-NP facility, Romania. The accuracy and reliability of
laboratory tests are determined by factors such as human factors; accommodation
and environmental conditions; methods of analyzing and validating methods; used
equipment; traceability of the measurement; sampling; manipulation of
measurement items. All of the factors above will contribute to total
measurement uncertainty. For this purpose an important role it has the
effective management of the technical aspects of the QMS.
Keywords
References
- [1]. Bhatt B. C. Radiat Prot Environ 34 (2011) 6-16.
- [2]. Tuncel N., Karayalcin B., Koca G. International Journal of Medical Physics, 2014; 2 (1)
- [3]. Rhynert C. R., Miller W. G. Health Phys.18 (1970) 681-4.
- [4]. Pradhan A. S., Lee J. I., Kim J. L. J Med Phys. 33 (2008) 85-99.
- [5]. Sahoo G. S., Tripathy S. P., Paul S., Sharma S. D., Sharma S. C., Joshi D. S., Bandyopadhyay T. J Med Phys. 39-4 (2014) 225-230.
- [6]. Deutsches Atomforum e.V., Berlin (Germany); Kerntechnische Gesellschaft e.V., Bonn (Germany); 1004 p; 2009; 4 p; Annual meeting on nuclear technology 2009; Jahrestagung Kerntechnik 2009; Dresden (Germany); 12-14 May 2009, A. Jahn, M. Sommer, J. Henniger, Radiation Protection and Dosimetry (S61).
- [7]. Stochioiu A., Luca A., Sahagia M., Margineanu R. M., Tudor I. Journal of Environmental Radioactivity 112 (2012) 4-7.
- [8]. ISO/IEC 17025:2005 - General requirements for the competence of testing and calibration laboratories.
Details
Primary Language
English
Subjects
-
Journal Section
Conference Paper
Authors
Dorina Aranghel
*
National Institute for Nuclear Physics and Engineering, Reactorului 30, RO-077125, POB-MG6 Bucharest-Magurele, Romania
Romania
Doina Stancıu
National Institute for Nuclear Physics and Engineering, Reactorului 30, RO-077125, POB-MG6 Bucharest-Magurele, Romania
Romania
Chivuta-ramona Badıta
National Institute for Nuclear Physics and Engineering, Reactorului 30, RO-077125, POB-MG6 Bucharest-Magurele, Romania
Romania
Ana Stochıoıu
İon Tudor
Publication Date
March 16, 2018
Submission Date
December 7, 2017
Acceptance Date
January 31, 2018
Published in Issue